Notes for electric installation on site
n
The fuse on the installation side must at least correspond to the max-
imum current consumption of the device (see type plate) and may not
exceed the value permitted for the mains plug.
n
Three-phase devices
Three-phase devices may only be operated on networks grounded
at the neutral point, e.g. TN-C, TN-C-S or TN-S. For ungrounded
networks (e.g. IT systems) or asymmetrically grounded networks, an
isolating transformer must be connected upstream and the neutral
point grounded. When selecting the isolating transformer, consider
the maximum connected loads of the device.
Ensure correct rotating field! If the rotating field is incorrect, an
error message is displayed and the device cannot be put into opera-
tion.
5.2 Switching on the device for the first time
You will find the entry sequence prescribed for safety reasons in the graphic.
The entries must be input each time the heat transfer liquid is changed and
after the device has been switched on for the first time.
Fig. 19: Asymmetrically grounded delta
network (without isolating transformer)
V4
Integral process thermostats with flow control unit
56 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...