4.4 LAUDA heat transfer liquids
DANGER!
Use of unsuitable heat transfer liquid
Fire
Select a heat transfer liquid with a temperature range suit-
able for the application.
CAUTION!
Use of unsuitable heat transfer liquid
Equipment damage due to freezing
Select a heat transfer liquid with a temperature range suit-
able for the application.
For general consideration:
The heat transfer liquid used must have an electrical conduc-
tivity greater than 1 µS/cm.
Silicone oils and mineral oils are not suitable for through-flow
control.
Observe the safety data sheet for the LAUDA heat transfer
liquid being used. If required, you can download corresponding
safety data sheets from our homepage.
Open the LAUDA homepage, tap
Services
Download
center.
In the Download center, chose the
[Safety data sheet]
option
in the
[Document type]
drop-down list.
A list of safety data sheets in PDF format in different lan-
guages is displayed.
Tap the relevant safety data sheet.
The download starts and the PDF file is downloaded.
Never use contaminated heat transfer liquid. Contamination in
the pump housing can cause the pump to seize and the device
to shut down.
For attention when using Kryo 30:
If the heat transfer liquid reaches the lower limit of the tem-
perature range, the temperature control properties can be
expected to deteriorate as a result of the increase in viscosity.
Therefore, only fully utilize this temperature range when abso-
lutely necessary.
The water content decreases during longer periods of operation
at higher temperatures and the mixture becomes combustible
(flash point 119°C). Check the mixing ratio using a hydrometer.
V4
Integral process thermostats with flow control unit
33 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...