ID Function
Unit
Command
57 Control parameter XpF
[–]
IN_PAR_10
61 Control parameter Prop_E
[K]
IN_PAR_15
Table 14: Control
ID Function
Unit
Command
59 Set point offset
[K]
IN_PAR_14
67 Control in control variable X: 0 = internal / 1 = external Pt / 2 = external
analog / 3 = external serial / 5 = external Ethernet / 6 = external EtherCAT /
7 = external Pt 2 (only for Integral)
[–]
IN_MODE_01
69 Offset source X for set point: 0 = normal / 1 = external Pt / 2 = external
analog / 3 = external serial / 5 = external Ethernet / 6 = external EtherCAT /
7 = external Pt 2 (only for Integral)
[–]
IN_MODE_04
Table 15: Rights
ID Function
Unit
Command
63 Status of Master keyboard: 0 = free / 1 = blocked
[–]
IN_MODE_00
65 Status of remote control keyboard: 0 = free / 1 = blocked
[–]
IN_MODE_03
Table 16: Status
ID Function
Unit
Command
75 Status of standby: 0 = Device is switched on / 1 = Device is switched off
[–]
IN_MODE_02
107 Device type (e.g.: "INT" or "VC")
TYPE
130 Device status: 0 = OK / -1 = fault
[–]
STATUS
131 Fault diagnosis; a 7-digit answer in the format XXXXXXX is output, whereby
each character X contains fault information (0 = no fault / 1 = fault).
The following information is defined for the seven places of the answer
format:
n
1st character = Error
n
2nd character = Alarm
n
3rd character = Warning
n
4th character = Overtemperature
n
5th character = Low level
n
6th character = 0 (with alarm setting: High level
n
7th character = External control value missing
[–]
STAT
V4
Integral process thermostats with flow control unit
48 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...