6.16
Programmer
6.16.1
Basic information
The programmer allows you to perform and save a temperature-time pro-
gram. A program consists of several temperature-time segments A segment
contains information on the final temperature of the segment, the duration,
the temperature tolerance, the pump level and the switch position (off/on)
of the contactmodule. Ramps, temperature jumps and temperature mainte-
nance phases are possible.
n
Ramp
A ramp is defined by the specified duration between the start and
the end of the segment, and by the destination temperature, i.e. the
temperature at the end of the segment.
n
Temperature jump
If a time is not specified (time is equal to 0), the end temperature is
reached as quickly as possible.
n
Temperature maintenance phase
No temperature change (i.e. the temperatures at the start and the end
of a segment remain identical).
n
Pump level off or 0
Pump level [---] (means pump is off) can be selected within a
segment. As a result, the program ends when this segment is
reached, even though other segments follow in the program. The
status of the thermostat is set to “Standby”. When the program is
started, a message appears indicating that the program will end at
this segment with pump level 0.
n
Program optimization
Activating program optimization yields a very good control action in
practice. With programs including both ramps and other types of seg-
ments, the actual temperature profile matches the target temperature
profile more closely than programs without optimization. It reduces
overshoots. There can only be large undershoots at the ramp ends if the
control parameters are very unfavorable. Deactivate optimization in this
case.
Tolerances
that are too small will impair the control result. Work without
tolerances where possible.
n
Standby
If the device switches to standby while a program is running, the active
program is automatically paused.
1.
Press the Enter key to open the menu.
2.
Select the
Programmer
Program X
menu item.
The submenu in the selected program opens.
Fig. 63: Programmer
V4
Integral process thermostats with flow control unit
104 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...