1.
Start the Windows command processor by entering
cmd.exe
@
on
the PC with Microsoft Windows operating system.
The entry window will open.
2.
There are two ways of checking:
n
Enter the ping command together with the IP address.
ping XXX.XXX.XXX.XXX
@
For "XXX.XXX.XXX.XXX", put the IP address that was entered
when the Ethernet interface was configured.
Or
n
Enter the ping command together with the serial number of the
thermostatic circulator (possible with software control system
1.36 or later).
ping serial number
@
If the Ethernet interface is configured and connected correctly,
the interface will return four responses within a very short time.
Fig. 15: Example for entering the ping command
It is easy to check the connection to the interface with a PC and Microsoft
Windows operating system.
n
For Windows 3.11 with the program "Terminal".
n
For Windows 95/98/NT/XP the program "HyperTerminal".
n
For operating systems Windows Vista, Windows 7, Windows 8 and
Windows 10 "HyperTerminal"* is not part of the operating system.
Checking the LAN network
Check the LAN network and the process
interface
V4
Integral process thermostats with flow control unit
41 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...