WARNING!
Risk of heat transfer liquid overheating due to incorrect entry of
overtemperature switch-off point Tmax
Fire
Set the overtemperature switch-off point to 5 K above the
upper limit of the temperature range for your application.
The overtemperature switch-off point must be below the
flash point of the heat transfer liquid.
WARNING!
Splashing heat transfer liquid
Eye damage
Always wear suitable safety glasses when working on the
device.
WARNING!
Overflow of heat transfer liquid
Electric shock
Do not overfill the device. Observe the level display and the
thermal volume expansion of the heat transfer liquid.
WARNING!
Spraying of heat transfer liquid
Electric shock
Do not spray heat transfer liquid. Use a funnel for filling.
WARNING!
Overflow of heat transfer liquid due to increase in volume
caused by heating
Scalding, electric shock
Take into account the increase in volume caused by heating
of the heat transfer liquid.
CAUTION!
Risk of heat transfer liquid escaping
Slipping hazard
Draining tap must be closed.
V4
Integral process thermostats with flow control unit
71 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...