ID Function
Unit
Command
52 Control parameter TdE
[s]
OUT_PAR_07_XXXX.X
54 Correction limitation
[K]
OUT_PAR_09_XXX.X
56 Control parameter XpF
[–]
OUT_PAR_10_XX.X
60 Control parameter Prop_E
[K]
OUT_PAR_15_XXX
Table 25: Control
ID Function
Unit
Command
58 Set point offset
[K]
OUT_PAR_14_XXX.X
66 Control in control variable X: 0 = internal / 1 = external Pt /
2 = external analog / 3 = external serial / 5 = external Ethernet /
6 = external EtherCAT / 7 = external Pt 2 (only for Integral)
[–]
OUT_MODE_01_X
68 Offset source X for set point: 0 = normal / 1 = external Pt /
2 = external analog / 3 = external serial / 5 = external Ethernet /
6 = external EtherCAT / 7 = external Pt 2
[–]
OUT_MODE_04_X
Note (ID 66 and 68): If X = 3, the commands ID 66 and ID 68 cannot be executed in some constant temperature control
devices until an external temperature specification has been received (via the command ID 15).
Table 26: Rights
ID Function
Unit
Command
62 Master keyboard (equivalent to "KEY"): 0 = unlock / 1 = lock
[–]
OUT_MODE_00_X
64 Keyboard remote control unit (command): 0 = unlock / 1 = lock
[–]
OUT_MODE_03_X
Table 27: Status
ID Function
Unit
Command
74 Switch the device on / off (standby)
[–]
START / STOP
Table 28: Programmer
ID Function
Unit
Command
76 Select the program for the subsequent commands (X = 1 – 5). The
default program is 5 when the constant temperature equipment is
switched on.
Caution!
Executing this command may stop any programs that are currently
running.
[–]
RMP_SELECT_X
78 Start programmer
[–]
RMP_START
79 Pause programmer
[–]
RMP_PAUSE
V4
Integral process thermostats with flow control unit
52 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...