7.7 Checking the low-level protection
An alarm signal sounds before the liquid level decreases so much that the
heater is no longer covered completely with liquid.
Low level
appears on the
display. The electronics system switches off the device components.
WARNING!
Contact with hot or cold heat transfer liquid
Scalding, cold burns
Bring the heat transfer liquid to room temperature before
draining.
An alarm must sound as soon as the minimum level is reached.
Protective equipment:
n
Safety glasses
n
Protective gloves
n
Protective work clothing
1.
Switch on the device and the pump. Set the set temperature to room
temperature.
2.
Watch the level indication on the display.
3.
Lower the liquid level in the device. For this purpose, allow the heat
transfer liquid to flow out into a suitable container via the open drain
nozzle.
The display shows that the heat transfer liquid is sinking.
If the fill level falls to level 2, a warning is issued on the display.
If the liquid falls to level 0, the device switches off and the
Alarm
message appears on the display.
4.
Close the drain tap.
5.
Top up the heat transfer liquid.
The liquid level in the display rises.
6.
Delete the fault in the device menu.
7.8 Check the heat transfer liquid
Protective equipment:
n
Safety glasses
n
Protective gloves
n
Protective work clothing
V4
Integral process thermostats with flow control unit
131 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...