Personnel:
n
Operating personnel
1.
Place the device on a suitable level surface.
The devices can be moved. To do this, release the parking
brakes on the castors by pushing the lever upward.
2.
Lock the castors of the device. To lock, press down the lever on the
castor.
4.2 Hoses
CAUTION!
Risk of external hydraulic circuit bursting
Scalding, cold burns
Use hoses with a greater compressive strength than the
maximum possible pump pressure.
CAUTION!
Risk of heat transfer liquid escaping due to the use of unsuitable
hoses
Scalding, cold burns
The temperature and media resistance of the hoses must be
suitable for the application.
CAUTION!
Contact with hot or cold hoses
Hot and cold burns
Use insulated hoses for temperatures below 0 °C and
above 70 °C.
Please note:
n
The thread connections of the pump connectors or the thread connec-
tions of the screw cap and the seat of the nut must be moistened with
lubricant.
n
When laying the water cooling and heat transfer liquid hoses, make sure
that the hoses cannot be kinked or crushed.
V4
Integral process thermostats with flow control unit
28 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...