n
[Control parameters]
This data record is located in the directory
LAUDA / USER
and contains
the file
CTRLPARA.INI
. Only such a file can be exported. The file will
be overwritten on re-export.
The file contains all the temperature control parameters as well as the
temperature limits Tih and Til. The file is designed to transfer the settings
of the control parameters from one device to another.
n
[Analog module config.]
[Contact module config.]
These data records can only be exported if an analog module or con-
tact module has been connected to the constant temperature equip-
ment. The configuration is stored in the LAUDA / USER directory in
the
ANACFG.INI
or
CONTCFG.INI
file. Only such a file can be
exported in each case. This file will be overwritten on re-export. The file
contains the complete configuration of the connected analog module or
contact module. The file is designed to transfer the configuration from
one constant temperature device to another.
Personnel:
n
Operating personnel
1.
Insert the USB stick in the USB host on the operating unit of the
machine.
2.
Press the Enter key to open the menu.
3.
Select the
Data export/import
Data export
menu items in the
main menu.
The submenu opens.
4.
Select one of the following options:
n
n [Temperature graphic]
n
n [Device status]
n
n [Device data]
n
n [Control parameters]
n
[Analog module config.]
n
[Contact module config.]
5.
Start the export by pressing the Enter key [OK].
Messages appear on the display during the export:
n
Data export to USB stick started.
n
Data export to USB stick has been successfully completed.
Confirm the message by pressing the Enter key [OK] once the
data export has been completed.
If the data export was not successful, the message "Data export to USB stick
failed" appears on the display. Press [OK] to confirm.
Check that the USB stick was inserted correctly and that there is enough
free memory space (at least 1 MB).
Start the data export again.
V4
Integral process thermostats with flow control unit
125 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...