Each program starts with the segment
Start
, which determines the tem-
perature at which segment
1
continues the program. The temperature of the
Start
segment is reached as quickly as possible. It is not possible to specify
a time limit for the
Start
segment. Without the
Start
segment, segment
1
would be different depending on the heat transfer liquid temperature at
program start.
Table 33: Program example, before (
_____
)
No.
Tend
hh
:mm
Tolerance
Pump
S1
S2
S3
Start
30.00
---
---
0.0
---
off
off
off
1
30.00
0
20
0.1
2
off
off
off
2
50.00
0
20
0.0
3
off
off
off
3
70.00
0
40
0.0
4
off
off
off
4
70.00
0
10
0.1
2
off
off
off
5
60.00
0
30
0.0
2
off
off
off
6
40.00
0
0
0.0
2
off
off
off
In the edited table, a new segment with the number 3 has been entered.
The time and the pump level for segment 4 have also been modified. The
tolerance and pump level for segment number 5 have been adapted.
Table 34: Program example, after (- - - - dashed line, edited)
No.
Tend
hh
:mm
Tolerance
Pump
S1
S2
S3
Start
30.00
---
---
0.0
---
off
off
off
1
30.00
0
20
0.1
2
off
off
off
2
50.00
0
20
0.0
2
off
off
off
3
50.00
0
20
0.1
3
off
off
off
4
70.00
0
20
0.0
4
off
off
off
5
70.00
0
10
0.8
2
off
off
off
6
60.00
0
30
0.0
2
off
off
off
7
30.00
0
0
0.0
2
off
off
off
In the Integral T in the Pumpcolumn
[off]
or
[on]
can be entered.
Note the following and see Fig. 67:
n
The Tolerance field ensures strict compliance with the residence time at
a specific temperature, for example.
n
The subsequent segment is only processed when the outflow tem-
perature reaches the tolerance range (1) so that the ramp in the second
segment is delayed and only starts at 2, for example.
Start segment
Tolerance
V4
Integral process thermostats with flow control unit
107 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...