In the basic window of the constant temperature equipment, the pressure of
the heat transfer liquid at the outlet of the flow controller is displayed in the
direction of the application. Next to it the corresponding measured flow in
the direction of the application.
Please note:
n
The system, consisting of constant temperature equipment, application
and connecting hoses, must be properly filled, vented and degassed
n
Once the through-flow control has been activated (
Status on
), the
pump power is automatically controlled by the flow controller.
n
The application (for example, glass apparatus) and all hoses used must be
suitable for the resulting pressure.
If the adjustable maximum pressure is exceeded, the constant tem-
perature equipment stops and reports the
gage pressure
alarm.
n
Install a separate safety valve against gage pressure in the circuit to
protect against operating errors.
Fig. 42: Control menu in the Integral XT for the flow controller
Fig. 41: Basic windows Integral XT and Inte-
gral P with through-flow control
V4
Integral process thermostats with flow control unit
78 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...