Lake Shore Model 340 Temperature Controller User’s Manual
10-6
Options and Accessories
HOLE #4
HOLE #3
HOLE #1
HOLE #2
IGNORE
ANALOG OPTION PLUG
TRANSFORMER ASSEMBLY
ARROW
NUB
PLUG
SLOT
Figure 10-6. Model 340 Mounting Holes and Analog Option Plug
6. With the instrument still standing on its face, turn it to view the inside circuit board. The upper half of the
circuit board above the transformer assembly contains four holes where metal standoffs install.
Find Hole #1 to the left between C245 and C265.
Find Hole #2 4.5 inches to the right of Hole #1, and just to the left of C298 and U154.
Find Hole #3 4.5 inches to the right of Hole #2, and just to the left and below the Digital Option plug (J9).
Find Hole #4 2.5 inches above Hole #3, just to the left and above the Digital Option plug (J9). Ignore hole
to left and beneath Hole #4.
Into each hole, insert a 4-40 screw through the bottom of the circuit board, and thread it into a metal
standoff on the top of the circuit board.
HOLE #4
HOLE #1
HOLE #2
INPUTS
HOLE #3
ANALOG
PLUG
Figure 10-7. Mounting Holes & Analog Plug
7. Just beneath the transformer assembly is the Analog Option plug (J18). Plug either end of the ribbon
cable into the Analog Option plug (J18). Slide the arrow nub down the plug slot for correct positioning.
8. Plug the other end of the ribbon cable into the 3462 Option Card plug (J3) located on the lower right
corner of the card.
9. Place the 3462 Option Card on top of the metal standoffs in the instrument. Align the holes in the card
with the corresponding metal standoffs. Fit the 3462 sensor inputs out the rear panel option opening.
Insert a 4-40 screw through each 3462 Option Card hole and thread it into the metal standoff beneath.