
Lake Shore Model 340 Temperature Controller User’s Manual
Remote Operation
9-37
MDAT?
Query Min/Max Data for an Input
Input: MDAT?
<input>
Returned: <min value>,<max value>
. Format:
±
nnn.nnnE
±
n,
±
nnn.nnnE
±
n[term]
Remarks:
Returns the minimum and maximum input data.
<input>
specifies which input to query.
MDATST?
Query Min/Max Data Status for an Input
Input: MDATST?
<input>
Returned: <min bit weighting>, <max bit weighting>
. Format:
nnn,nnn[term]
Remarks:
The integer returned represents the sum of the bit weighting of the input minimum /maximum
data status flag bits.
<input>
specifies which input to query. See
RDGST?
Command.
MNMX
Configure Minimum and Maximum Input Function Parameters
Input:
MNMX <input>, [<on/pause>], [<source>]
Returned:
Nothing
Remarks:
Configures the minimum and maximum input functions.
<input>
Specifies input to configure.
<on/pause>
Turns on or Pauses the max/min function. Valid entries:
1
= on,
2
= paused.
<source>
Specifies input data to process through max/min. Valid entries:
1
= kelvin,
2
= Celsius,
3
= sensor units,
4
= linear data.
Example: MNMX B, 1, 3[term] -
Input B min/max function is on and processes data from the input
sensor units reading.
MNMX?
Query Minimum and Maximum Input Function Parameters
Input: MNMX?
<input>
Returned: <on/pause>, <source>
. Format:
n,n[term]
Remarks:
Returns a min/max input configuration. See the
MNMX
command for parameter descriptions.
<input>
specifies which input to query.
MNMXRST
Resets Min/Max Function for All Inputs
Input: MNMXRST
Returned:
Nothing
Remarks:
Resets the minimum and maximum data for all inputs.
MODE
Configure Remote Interface Mode
Input: MODE
<mode>
Returned:
Nothing
Remarks:
Configures the remote interface mode.
<mode>
specifies which mode to operate. Valid
entries:
1
= local,
2
= remote,
3
= remote with local lockout.
Example: MODE 2[term] -
Places the Model 340 into remote mode.
MODE?
Query Remote Interface Mode
Input: MODE?
Returned: <mode>
. Format:
n[term]
Remarks:
Returns the remote interface mode. See the
MODE
command for parameter descriptions.
MOUT
Configure Control Loop Manual Output Value
Input: MOUT
<loop>,
<value>
Returned:
Nothing
Remarks:
Configures the control loop manual output values.
<loop> s
pecifies loop to configure.
<value>
specifies value for manual output.
Example: MOUT 1, 22.45[term] -
Control Loop 1 manual output is 22.45%.