128
CHAPTER 6:
Maintenance and Troubleshooting
Model CRX-6.5K Probe Station
6.2.2 Cleaning the
Vacuum Chamber
Exterior
The exterior of the probe station chamber should be kept generally clean and clear of
dust and other possible contaminants. Surfaces may be wiped down using damp, lint-
free cloths like Kimwipes®. Isopropyl alcohol on a lint-free cloth is recommended to
loosen adhesives and other residue. Dust on the exterior may also be removed with a
commercially available compressed air product such as Dust Off®.
6.2.3 Cleaning the
Vacuum Chamber
Interior
Failure to keep the chamber clean and store it under vacuum will result in poor probe
station performance due to contaminants and moisture in the chamber. A dirty
chamber requires longer pump down times, more helium to cool, and results in
higher base temperatures. If the Model CRX-6.5K chamber is not properly
maintained, it will be increasingly difficult to operate the sample stage at the
specified base temperature.
1. Prior to cleaning, remove the following:
D
Vacuum chamber lid
D
Radiation shield lid
D
Main chamber o-ring
D
Sample holder
D
Probes
2. Wipe down the following surfaces with a lint-free wipe like Kimwipes® and
isopropyl alcohol. Do not clean the viewports during this step.
D
Top edge of the radiation shield (mating surface)
D
Mating surfaces of the radiation shield lid
D
Any surface that shows fingerprints
D
Any surface likely touched during sample exchange
D
Sealing surface of the vacuum chamber lid
D
Main chamber o-ring
D
Main chamber o-ring groove
The radiation curtains are very fragile; do not soak them or rub them as they can
come off or break.
3. Clean and lubricate the chamber components:
D
Clean the viewport windows (section 6.2.4)
D
Lubricate the main chamber o-rings with a thin layer of high quality vacuum
D
Clean the sample holder (section 6.2.6)
4. Reassemble the probe station:
D
Sample holder
D
Probes
D
Main chamber o-ring
D
Radiation shield lid
D
Vacuum chamber lid
5. Evacuate the chamber for 30 min prior to loading a sample.
To minimize the risk of contamination, the radiation shield lid and vacuum chamber
lid should remain in place except when working in the chamber. Lake Shore also
recommends that you store the chamber under vacuum to reduce oxidation.
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...