76
CHAPTER 3:
Installation and Setup
Model CRX-6.5K Probe Station
5. As you slide the microwave dowel onto the probe arm, note the relative height of
the microwave connector socket on the probe and the microwave connector plug
on the semirigid cable. If the relative heights of the two connectors do not match,
the semirigid cable can be gently bent using the thumb and forefinger to align
the connector plug with the height of the connector socket (FIGURE 3-37).
6. Continue to slide the microwave dowel onto the probe arm until the microwave
connector socket aligns with and contacts the connector plug on the semirigid
cable. The semirigid cable can be gently lifted or pushed to one side or the other
to help align the two connectors.
7. Carefully start threading the connector plug onto the probe connector socket.
The probe can be rotated slightly to better align it as you thread. Refer to
section 5.3.6.3 if the cable length seems inappropriate.
8. If the connector is properly aligned, it will make good contact simply by finger
tightening. It is acceptable if the back of the copper probe body is not flush with
the copper end of the probe arm.
A wrench should not be used to tighten the plug onto the microwave probe unless it is
a calibrated torque wrench specifically designed for making microwave connections.
9. Once the microwave connector is tight, use the 1.5 mm hex driver to tighten both
probe arm set screws. If the probe body was rotated in order to ease the
alignment of the microwave connectors when threading, hold the microwave
probe body with your thumb and forefinger as you tighten the set screws to
vertically align the probe.
FIGURE 3-37
Left: Misaligned probe; Right: A properly aligned probe
FIGURE 3-38
Microwave probe threaded onto a semirigid cable
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...