3.7.3.2 Removing the ZN50 Probe Mount
1. If you have not yet removed the blade, you will need to disconnect the SMA
connector. The strain relief to the SMA connector (on the cryogenic coaxial cable)
must be held in place with tweezers or your fingers while removing the SMA plug.
2. Using the x-axis hand dial and the y-axis micrometer, move the probe arm to
expose the braid block mounting holes in the sample stage.
3. Using the 2.5 mm hex driver, loosen the two captive screws on the braid block
4. Pull the braid block out.
5. Using the 1.5 mm hex driver, loosen both M3 probe arm set screws two to three
turns and slide the dowel out of the probe arm (FIGURE 3-31).
6. Pull the probe mount out of the probe station.
3.7.4 Installing a
Microwave Probe
The K-connector, 2.4 mm connector and the V-connector all look very similar;
however, they may not mate to a connector of different frequency rating and the
probe or cable may be damaged if forced to thread to an improper mate. See
section 2.5.1 for more information. Contact Lake Shore if you are still unsure of your
probe station’s microwave configuration.
Unlike the ZN50 blade and optical fiber assemblies that use a separate probe mount,
the microwave probe has an integrated mount that attaches to the probe arm, and
includes probe mount braids and a braid block for attachment to the thermal
anchor point.
GSG microwave probe tips are extremely delicate; the slightest touch on the probe tips
can disturb the ground-signal-ground (GSG) transmission line geometry, thereby
degrading performance. Due to the high cost of replacement microwave probes, we
highly recommend that users become comfortable with probing techniques using ZN50
probes prior to installing and probing with microwave probes to avoid damage to the
microwave probe tips (Lake Shore provides two ZN50-25-BECU and two ZN50-25-W
probe tips in the probe starter kit).
Follow this procedure to install a microwave probe. If a ZN50 or optical fiber probe
mount is installed in your probe station, you will need to remove it first
(section 3.7.3.2 or section 5.3.7.1).
1. Follow steps 1 to 3 in section 3.7.1.
2. Remove a microwave probe from its storage case.
3. Grasp the microwave probe between your thumb and forefinger on the sides of
the probe body so that the dowel is extending outward from your hand, and the
braid block is dangling below your hand (FIGURE 3-36).
4. Using the 1.5 mm hex driver, loosen the set screws on the probe arm and slide the
dowel of the microwave probe into the probe arm (FIGURE 3-31).
FIGURE 3-36
Proper handling of a microwave probe
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...