70
CHAPTER 3:
Installation and Setup
Model CRX-6.5K Probe Station
4. Connect the bucket between the turbo pumping system and the probe station’s
vacuum isolation valve.
3.7 Installing and
Removing Probes
Probes are installed at the ends of probe arms, inside the chamber. The chamber and
radiation shield must be opened. It may also be helpful to remove the sample holder
to make more room to work inside the chamber. Four ZN50 probes are included in the
probe starter kit for training purposes.
The probes are packaged separately to protect the delicate tips. Do not touch the tips. Do
not handle the alumina blade or the electrical conductors on the ZN50 probe with bare
hands, as this may reduce its isolation. Wear nitrile gloves while changing blades.
3.7.1 Installing a Probe:
Prep Instructions for All
Probe Types
Follow this procedure to prepare for installation of all probe types, then use the
unique instructions for each individual probe type.
1. Open the vacuum chamber and radiation shield using the guidelines in
2. If you purchased semirigid cables or coaxial cables, there will be safety ties on the
arms. Cut these off, and remove the ties from all cables. Be careful not to cut the
Kapton® tape on the coaxial cables. You may need to use the x-axis hand dial to
move the probe arms in or out so that the ties are more accessible.
3. Using the x-axis hand dial, extend the probe arm into the chamber until the probe
arm set screws are accessible.
3.7.2 Installing a
ZN50 Probe
The ZN50 series probes consist of a probe mount, ceramic blade with SMA electrical
connector, and probe tip. Use the procedure in section 3.7.2.1 to install a ZN50 probe.
If the probe mounts are already installed (FIGURE 3-28), go to section 3.7.2.2 and
begin with installing the probe blade. If the probe mount is not installed, please be
aware that some probe station users find it easier to first install the blade to the probe
mount outside the station (section 3.7.2.2), and then install the probe to the
probe arm.
FIGURE 3-27
Pump line vibration isolator
SMA
connector
mount
blade
tip
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...