46
CHAPTER 3:
Installation and Setup
Model CRX-6.5K Probe Station
3.2.3 Power
Requirements
Electrical power is required for the operation of the Model 336 temperature
controller, vision system, turbo pumping system, and CCR compressor. The Model
CRX-6.5K requires 1-phase power only. Most equipment is designed to operate over a
range of line voltages. Some equipment must be configured to operate at a specific
voltage within the range listed. This equipment is configured at Lake Shore to the
voltage specified when the equipment is ordered. Refer to section 3.4 for making
utility connections and section 6.4.1 for additional information on power
requirements and configuration options.
The electrical equipment can be grouped as shown in TABLE 3-1 and TABLE 3-2 to
distribute the system power requirements over multiple facility circuits. TABLE 3-1
details the operational power requirements for each circuit, and TABLE 3-2 details
the operational power requirements for the CCR compressor. The customer must
provide wiring (that follows local safety codes) to the CCR compressor.
The turbo pumping station should be powered from a separate circuit to isolate
electrical noise from the sensitive electronic instruments. The circuit protection used
for the turbo pumping system should be capable of handling high inrush currents.
The CCR compressor will draw high inrush currents when powered on. Circuit
protection capable of handling these short duration inrush currents is required to
prevent nuisance tripping of the safety devices. Thermal circuit breakers perform best
when available. The necessary utility connections are described in section 3.4.
Probe station equipment
Operational current required
100 VAC
120 VAC
220 VAC
240 VAC
Circuit 1
Model 336 temperature controller
2.5
2.1
1.2
1.1
Vision system
(monitor, camera and light source)
7.2
6.0
3.3
3.0
Circuit 2
TPS-FRG turbo pumping station
2.6
2.2
1.2
1.1
TABLE 3-1
1
-phase power requirements
Probe station
equipment
Power line voltage (±10%)
Operating current
Starting current
Recommended
circuit breaker
Power requirement
Circuit 3
HC4E1
1 phase,
208/230 V~ (± 5%), 60 Hz
200 V~ (± 5%), 50 Hz
230/240 V~ (± 5%), 50 Hz
Max 13/16 A
(50/60 Hz)
68 A
(50/60 Hz)
20 A
3.0 kW at 208 V, 60 Hz
2.6 kW at 200 V, 50 Hz
TABLE 3-2
1-phase power requirement for CCR
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...