93
|
www.lakeshore.com
does happen, please refer to section 6.2.12). It is possible that the sample cooling
assembly may suffer irreparable damage.
Never operate the probe station above room temperature without the CCR on and the
probe station under vacuum.
When operating the sample stage above room temperature, it is best to leave the
vacuum isolation valve open and the pump running during operation. This is because
at elevated temperatures the system will slowly outgas and degrade the vacuum. The
pump line isolation option (PS-PLVI-40) should be used if measurements are
routinely made at elevated temperatures to reduce vibrations due to the vacuum
pump.
4.4.1.2 Vacuum Gauge Location
For convenience, the vacuum gauge on the PS-V81DPC pump option is located on an
NW 25 T immediately at the inlet of the turbo pump. Vacuum pressures in the probe
station chamber will be a half order of a magnitude higher than measured here. All
probe station vacuum specifications are made with the gauge located on the vacuum
chamber itself. For more accurate vacuum pressure readings the gauge can be moved
to an available port on the vacuum chamber or as shown in FIGURE 6-4.
4.4.1.3 Vacuum Performance
A room temperature vacuum level of <10
-3
Torr in the vacuum chamber is required for
the probe station to meet its specified base temperature and cooling time. Each probe
station is tested prior to shipment to ensure it can achieve appropriate vacuum level
and cooling specifications with a PS-V81DPC turbo pump option.
If the system will not reach a vacuum level of <10
-3
Torr within 1 h, or it takes more
than 10 min for the turbo pump to reach full speed, suspect a leak or other problem
with the vacuum system and refer to section 6.3.1.8 for troubleshooting information.
Even with properly functioning equipment and good maintenance, vacuum
performance of the probe station may degrade with time. The probe station may
need to be warmed up and re-evacuated if it has been in continuous operation for
several days or if the sample is warmed and cooled frequently.
To keep the vacuum system operating efficiently, refer to section 6.2.1 for a
preventive maintenance schedule.
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...