90
CHAPTER 4:
Basic Operation
Model CRX-6.5K Probe Station
smaller samples, respectively. Silver paint may be air dried or baked
according to manufacturer’s recommendations.
D
To remove: you can remove the silver paint by soaking it in acetone.
4.3.3.5 Permanent Mounting
Permanent mounting is generally considered a last resort if all other methods have
failed to give adequate performance. These methods generally do permanent
damage to the sample, sample holder or both. Low temperature solder and filled
epoxy (Stycast®) are permanent mounting options that are compatible with vacuum.
4.3.4 Mounting the
Sample Holder onto the
Sample Stage
Follow this procedure to mount the sample holder onto the sample stage.
1.
Raise
each probe 3 mm to 4 mm above the sample holder using the z-axis
micrometers.
2.
Center
all probes using the y-axis micrometers.
3.
Retract
all probes away from the sample stage using the x-axis hand dials. This
will provide maximum access to the sample stage.
4. Lower the sample holder onto the sample stage using the lifter tool (FIGURE 4-6).
5. If you are mounting an optional triaxial or coaxial sample holder, align the socket
on the sample holder with the cable location.
6. Using the 2.5 mm hex driver, fasten the sample holder to the sample stage by
starting all four M3 screws in a few threads before tightening any screws to pre-
vent cross-threading.
7. Tighten the four screws securely; this is the source of thermal contact between
the sample holder and the sample stage. It is best to tighten all four screws a little
at a time until snug.
Do not over-tighten, as damage to the sample stage threads is costly to repair.
8. If you are mounting an optional triaxial or coaxial sample holder, connect the sig-
nal cable after mounting the sample holder to the sample stage (FIGURE 4-6).
Using tweezers, line up the white dots and push the cable plug into the sample
holder socket.
Be very careful that the tweezers do not slip off the cable plug and onto the wire,
where they can accidentally pull the wire out of the connector.
9. If you are using microwave probes, make sure translation and planarization are
within range (section 4.6.3).
FIGURE 4-6
Left: Fastening the sample holder onto the sample stage; Right: Connecting the signal cable on a triaxial or coaxial sample holder
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...