12. Remove anything temporarily fastening the cable to the end of the probe arm.
Take care not to remove the tape that electrically insulates the SMA connector
(FIGURE 5-15).
5.3.2 Removing a
Micro-manipulated
Translation Stage
(MMS-07)
This section provides directions for removing a micro-manipulated translation stage.
Before removing the micro-manipulated translation stage, find a small tray in which
to put the hardware removed from the system.
1. Using the x, y and z-axes, center the probe arm in the chamber.
2. To prevent damage to a ZN50 or microwave probe, remove it from the probe arm
using the instructions in section 3.7.3 and section 3.7.5 respectively. Do not
remove optical fiber probes; they should be left in place.
3. Use a 7 mm wrench to loosen the hex screws that hold the spade lugs at the ends
of the arm shield braids to the radiation shield (FIGURE 5-9), but do not fully
remove these screws. Slide the spade lugs from under the screw heads and let the
arm shield braids fall.
4. Using a 3 mm hex driver, remove the four M4 screws that attach the stage base to
the baseplate (FIGURE 5-10). If necessary, use the x-axis hand dial to move the
stage forward to access all of the screws.
5. Using a 3 mm hex driver, detach the bellows from the chamber by removing the
FIGURE 5-8
Left: Slide the spade lug under the hex screw head and washer; Right: Tighten the hex screw
FIGURE 5-9
Left and Right: Use a wrench to remove the arm shield braids from the radiation shield
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...