CHAPTER 2:
System Overview
Model CRX-6.5K Probe Station
2.4.1.2 Avoiding Ground Loops
Ground loops are one of the most common noise sources in measurement systems
like a probe station. A ground loop occurs when two or more places on the probe
station are connected separately to ground. The loop area is exposed to magnetic
fields generated by AC power lines in the lab. The changing field induces line
frequency noise in the loop that can permeate through the measurement setup.
As mentioned in section 2.4.1.1, there are multiple places on the probe station that
connect electrically to the vacuum chamber, and therefore, have the potential to
form a ground loop. Most of these points are connected to provide high frequency
shielding for the measurement signals. The shield connections can be the source of
line frequency noise if they are allowed to create a ground loop.
Minimizing the effect of ground loops can be difficult—some experimentation may
be required to achieve the best results. One of the biggest difficulties is that signal
sources and acquisition electronics often operate with one lead internally referenced
to earth ground, so the overall grounding system must take this into account. The
best approach is to make the system a poor receiver for the noise:
Never remove earth safety ground protection from electronic equipment.
1. Electrically isolate any parts of the systems that do not require grounding for
safety or performance.
2. Attach cable shields at only one end of the cable if the shield conductor is not
being used to establish a return path for the signal.
3. Add resistance in series with reference ground leads in cases where some
common mode voltage is present.
FIGURE 2-20
DC/RF ground reference
Radiation shield stage
Probe
Sample stage
Sample holder
Cryogenic
coaxial cable
Signal connector
Probe
arm
base
Probe arm shield
To temperature
controller chassis
19-pin
connector
Feedthrough
Thermal
anchors
Shield
Vacuum chamber base
Содержание CRX-6.5K
Страница 4: ...Model CRX 6 5K Probe Station ...
Страница 52: ...42 CHAPTER 2 System Overview Model CRX 6 5K Probe Station ...
Страница 92: ...82 CHAPTER 3 Installation and Setup Model CRX 6 5K Probe Station ...
Страница 162: ...152 CHAPTER 6 Maintenance and Troubleshooting Model CRX 6 5K Probe Station ...
Страница 166: ...156 INDEX Model CRX 6 5K Probe Station ...