Tutorial
PCR-WE Interface Manual
327
TRIGger:SIMulation subsystem.
Trigger operation has three states: IDLE, INITiated, and WTG.
IDLE
INITiated
TRIG:SOURce?
WTG
Power line abnormality simulation start
Power line abnormality simulation end
ABOR
*RST
*RCL sent
INIT sent
BUS
TRG sent
IMM
To start a simulation immediately, set the trigger source to IMMediate, and then use
the INITiate command.
TRIGger:SIMulation:SOURce IMMediate
ʻ
Sets the trigger source to IMM
INITiate:SIMulation 'Initiates the SIMulation group. The power line abnormality
simulation begins.
To use software triggers to start the power line abnormality simulation on the SIMula-
tion group, change the trigger source to BUS.
TRIGger:SIMulation:SOURce BUS 'Sets the trigger source to BUS
INITiate:SIMulation 'Initiates the SIMulation group
TRIGger:SIMulation 'Applies a software trigger to the SIMulation group. The pow-
er line abnormality simulation begins.