[SOURce:]VOLTage Command
PCR-WE Interface Manual
191
VOLT:OFFS:LIM:UPP
Sets the upper DC voltage limit.
If the signal source is set to external signal source (VOLT:EXT:EXTDC:SIGN:SOUR
EXT), this command is invalid.
For single-phase three-wire, set this value with the phase voltage of phase U.
The following relationship must be met: VOLT:OFFS:LIM:LOW ≤ VOLT:OFFS|-
VOLT:OFFS:TRIG ≤ VOLT:OFFS:LIM:UPP
Command
[SOURce:]VOLTage:OFFSet:LIMit:UPPer <numeric>
[SOURce:]VOLTage:OFFSet:LIMit:UPPer?
Parameter
Value:
-445.0 to +445.0 Upper DC voltage limit (The default value is 445.0)
Unit:
V
Settings are reset to default values when an *RST command is sent.
Example
VOLT:OFFS:LIM:UPP -9
Response: NR3