MEASure/FETCh Command
PCR-WE Interface Manual
95
Measurement <meas-item> list of the total values of all phases
Phase designation (@chanlist) not possible
<meas-item>
Description
Unit
POWer[:DC]:TOTal
Total average power
W
POWer:AC[:REAL]:TOTal
Total power without a DC component
W
POWer:AC:APParent:TOTal
Total apparent power without a DC component VA
POWer:AC:REACtive:TOTal
Total reactive power without a DC component
VAR
POWer:AC:PFACtor:TOTal
Total power factor without a DC component
̶
POWer:ACDC[:REAL]:TOTal
Total power
W
POWer:ACDC:APParent:TOTal
Total apparent power
VA
POWer:ACDC:REACtive:TOTal Total reactive power
VAR
POWer:ACDC:PFACtor:TOTal
Total power factor
̶
Response
Returns the measurement of the designated phase in NR3[,<NR3>...] format in re-
sponse to FETC:<meas-item>? [(@chanlist)]/ MEAS:<meas-item>? [(@chanlist)].
Returns the measurement of the phase designated by the INST command in
NR3[,<NR3>...] format in response to FETC:<meas-item>?/ MEAS:<meas-item>?.
Returns 0 in response to a line voltage query during single-phase output.
Returns the measured values in the <meas-item> list order in comma-separated
NR3 format in response to FETC:ALL?/MEAS:ALL?. If you specify several phases,
the measured values of the first specified phase are returned in the <meas-item>
list order. Then, the measured values of the next specified phase are returned in
the <meas-item> list order.
Example
INST:NSEL 1;COUP NONE
FETC:CURR?
Returns the average current of the U phase.
INST:COUP ALL
FETC:CURR?
Returns the average currents of all phases.
FETC:CURR? (@2)
Returns the average current of the V phase.
FETC:CURR? (@1:3)
Returns the average currents of all phases.
FETC:ALL? (@1)
Returns all scalar measurements (25 items, excluding the total
value) of the U phase.
FETC:ALL?
Returns all scalar measurements (25 items 3, excluding the total
value) of all phases.