List of Errors
PCR-WE Interface Manual
309
Operation complete event errors
An error in the range [-899, -800] is used when the product wants to report an IEEE
488.2 operation complete event. This event occurs when the instrument
ʼ
s synchroni-
zation protocol, having been enabled by an *OPC command, completes all selected
pending operations.
The occurrence of any error in this class causes the Operation Complete bit (bit 0) in
the event status register to be set.
Error code
Error message description
-800 Operation complete
All selected pending operations in accordance with the IEEE
488.2, 12.5.2 synchronization protocol have completed.
Product-dependent errors
The occurrence of any error in this class causes the Device Dependent Error bit (bit 3)
in the event status register to be set.
■
Configuration change rejection errors
These errors occur when the specified configuration changes cannot be permitted.
Error code Error message
+100
Setting change denied while OUTPut ON state
+101
Setting change denied while TRANsient trigger in progress
+102
Setting change denied while SIMulation trigger in progress
+103
Setting change denied while PROGram trigger in progress
+104
Setting change denied while frequency synchronous.
+105
Setting change denied while BUSY
+106
Setting change denied while SIMulation in progress
+107
Setting change denied while PROGram in progress
+108
Setting change denied while SLEep mode
+109
Setting change denied while EXT PROGram SOURce selected
+110
Setting change denied while INT+EXT PROGram SOURce selected
+111
Setting change denied while V PROGram SOURce selected
+112
Setting change denied while SSTart in progress
+113
Setting change denied while SSTop in progress
■
Configuration conflict errors
These errors occur settings are in conflict with the existing settings.
Error code Error message
+201
Conflicts with OUTPut OFF state