Setting Up the Interface
PCR-WE Interface Manual
33
USB (standard)
To use the USB interface to control the product, a device driver that supports the USB
Test & Measurement class (USBTMC) must be installed on the controller. The USBT-
MC driver is installed automatically by the VISA library.
■
USB connection
Connect the product to a PC using a USB cable.
■
USB settings
The factory default USB setting is
“
USB enabled.
”
1
Press CONFIG (SHIFT+OPR MODE) > COM-I/F (F1) > USB (F2).
2
To enable USB, press ENABLE (F1). To disable it, press DISABLE (F2).
3
Press APPLY (F5).
To cancel, press CANCEL (F6).
4
Turn the PCR-WE
ʼ
s POWER switch off and then back on.
The settings take effect.
■
Service request
The product is equipped with service request and serial polling functions.
■
USB function
Complies with USB Specification 2.0
Complies with USBTMC Specification 1.0 and USBTMC-USB488 Specification 1.0
Baud rate: 480 Mbps maximum (high speed)
VID (vendor ID)
0x0B3E
PID (product ID)
PCR-WE: 0x104D
PCR-WE2: 0x104E
PCR-WE2R: 0x1054