4-24
PAN-A
4.3.5 Output ON/OFF control
This method is used to control the output's ON/OFF status using
external contact points. Output is turned off when the external contact
points are closed.
• The insulation of the external contact (S) and the
cable connecting it should be greater than the isola-
tion voltage of the unit. For details on the isolation
voltage of each model, see Chapter 7 "Speci
• When using shielded wires for connection, protect
the uncovered section of the shielded wire by using
insulation tubes with a withstand voltage greater than
the isolation voltage of the unit.
• Never attempt to carry out wiring to the control termi-
nals with the POWER switch turned ON. Otherwise,
electric shock or failure of the internal circuit may be
caused, and may result in improper operations.
• A negative voltage of approx. 0.6 V is generated
when the output is turned off, and a reverse current
of a few tens of mA may
fl
ow.
• In output of ON/OFF, "OFF" has higher priority. Therefore,
always set the OUTPUT switch to ON so that ON/OFF con-
trol using external contact points can be performed.
Connection and setup procedure
1. Turn off the OUTPUT switch.
2. Turn off the POWER switch.
3. Connect the contact points S (switch) across the control
terminals [10] and [11] as shown in Fig.4-16.
WARNING
CAUTION
NOTE