When the negative output terminal is
grounded
In summary, connect either output terminal to the chassis terminal for
safety reasons unless your application requires the output terminal to
be
fl
oating. Moreover, the output ripple may become large if the out-
put terminal is not connected to the chassis terminal. The speci
fi
ca-
tions of the power supply assume that the negative output terminal is
connected to the chassis terminal.
+
–
+
–
+
+S
–S
–
+
–
16
17
18
19
3
2
9
8
1
L
N
Load
Vext
Rext
PAN-A
Since the negative output terminal is at
ground potential, the section indicated in
gray must have an insulation capacity
that is greater than the maximum output
voltage of the power supply with respect
to the chassis.
GND
Approx. the
same potential
as the positive
output terminal
Control terminal
AC input
terminal
Output terminal
Short bar