6-4
PAN-A
Fig.6-2
Connection for voltage system adjustment
(PAN16-50A)
6. Turn on the POWER switch.
■
Warming-up
• To minimize adjustment error affected by initial drift, warm
up the unit at least 30 minutes before starting adjustment.
7. Set the output voltage to the rated output voltage.
For local control, turn the VOLTAGE control knob fully clock-
wise. For remote control, set the control signal to 10 V or 9.5 k
Ω
.
8. Turn on the OUTPUT switch.
9. Turn off the OUTPUT switch 30 minutes or more later.
Output voltage offset
10. Set the output voltage to 0 V.
For local control, turn the VOLTAGE control knob fully counter-
clockwise. For remote control, set the control signal to 0 V or 0
Ω
.
11. Turn on the OUTPUT switch.
12. Turn the CURRENT control knob clockwise until the unit
enters constant voltage operation. (Offset adjustments
must always be conducted in constant voltage operation.)
When the current is remote-controlled externally and the current
setting is 0 A, the unit may operate at a constant current, depend-
ing on the offset balance. To avoid this problem, turn the current
control knob so that a current equal to about 10 % of the rated
value can be provided at any time.
Rear OUTPUT
terminal board
Crimped
terminal
DVM
HI
LO
Connect the shorting bar
to the - (neg.) terminal.
NOTE