PAN-A
4-41
Chap
.4
Applied Oper
ation
1
Fig.4-24
Connection for master-slave-control series
operation (PAN16-50A)
• When carrying out master-slave-control series operation with
the PAN16-50A and PAN16-50 combined, refer to Fig.4-25
for connection of the CONTROL terminal board.
S
S
1
2
7
8
13
14
19
S
S
1
2
7
8
13
14
19
S
S
1
2
7
8
13
14
19
Master unit
rear OUTPUT
terminal board
Slave unit 1
Rear OUTPUT
terminal board
Slave unit 2
Rear OUTPUT
terminal board
Master unit
Rear CONTROL
terminal board
Slave unit 1
Rear CONTROL
terminal board
Slave unit 2
Rear CONTROL
terminal board
Connect any one of the + and - terminals
to the chassis ground terminal, using a shorting bar.
Load
[3]
[1]
[1]
[3]
Select the shortest
and heaviest gauge
wires practical for
this wiring
Select the shortest
and heaviest gauge
wires practical for
this wiring
NOTE