• Use a load cable of suf
fi
cient current capacity to
cover the rated value.
• After parallel connection, mount the output terminal
cover (on the front or rear panel).
To mount the front auxiliary output terminal cover,
see "3.4 Mounting the auxiliary output terminal
cover".
• Use wires of the same length and size for connecting
the load and each unit. Use of varying lengths and
size may cause different output current from each
unit.
• To assure good connection of load cables, use
crimped terminals.
• Position each unit with suf
fi
cient space around it. Do
not stack units a top one another.
• To minimize the in
fl
uence of noise on the output, use a 2-core
shielded wire or a twisted-pair wire to connect across the con-
trol terminals.
Note that longer lengths of wiring result in more susceptibil-
ity to the effects of noise; use of cables with anti-noise mea-
sures may not solve the problem.
• When using shielded wires, connect the shield between the
master and slave units to the + (pos.) output terminal of the
master unit. Connect the shield between the slave units to the
shield between the master and slave units.
• When performing remote sensing in master-slave-control par-
allel operation, prepare the sensing wiring only for the master
unit. For information on connection, see "4.2 Remote sens-
ing".
• If both parallel operation and remote sensing are required
when combining the PAN-A and PAN series, the error in the
current display of the PAN series used as slave units may
increase. In this case, use the slave units in sensing mode.