Instrument Messages
Event Messages
–226
List not same length
You are using the LIST structure, but have individual lists of differing lengths
–230
Data corrupt or
stale;
A legal data element was found, but it could not be used because the data
format or the data structure was not correct. This could occur if a new
measurement had been started but had not completed
–230
Data corrupt or
stale; Measurement
data is not available
Measurement data not available. The measurement that you are trying to get
data from must be the current active measurement. Maybe you have not
initiated the measurement, or it has not completed all the sweeps/averages
needed
–230
Data corrupt or
stale; Trace contains
no data.
Trace cannot be displayed because currently there is no data assigned to it.
Use the functions under the
Trace/Detector
menu, or load a previously
saved trace, to assign data to the trace
–230
Data corrupt or
stale; Unable to load
state from file
There is something wrong with the state data in the desired file. Maybe the file
is corrupt, or it is from an instrument/version that is not recognized by the
current instrument
–231
Data questionable
Indicates that the measurement accuracy is suspect
–232
Invalid format
A data element was found but it could not be used because the data format or
the data structure was not correct
–232
Invalid format; Map
information not
loaded
Instrument failed to load the burst mapping information from the selected file
–233
Invalid version
A legal data element was found but could not be used because the version of
the data is incorrect. For example, state data changes as new instrument
features are added, so old state files may not work in an instrument with a
newer version of software
–240
Hardware error
A legal program command or query could not be executed because of a
hardware error. The exact problem cannot be specifically identified
–240
Hardware error; See
details in Windows
Event Log under SA
The internal data acquisition system detected a problem at startup and logged
the details in the Windows Event Log
–241
Hardware missing
The operation could not be performed because of missing hardware; perhaps
the optional hardware is not installed
–241
Hardware missing;
Input not available
The hardware required is not part of this model or the option is not installed
–241
Hardware missing;
Internal preamp not
available at all
frequency points
The Internal Preamp is currently turned on, but the measurement is being
performed completely or partially outside the range of the preamp. It is
recommended that the user turns preamp off to ensure consistent results
across the entire measurement
–241
Hardware missing; not
available for this
model number
The hardware required is not part of this model
Err#
Message
Verbose/Correction Information
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...