96 Keysight N9038A MXE Service Guide
Instrument Messages
Event Messages
–200
Execution error; No
ranges are defined.
Activate a range.
There are no active ranges in the range table. You will need to activate at least
one range
–200
Execution Error;
Preselector centering
failed
Algorithm failed to center the preselector. This maybe caused by the signal
peak being too low in amplitude, or it could be due to excessive CW input
signal, alignment error, or hardware failure
–200
Execution Error;
Signal not stable
enough to track
The signal that you have selected to track is changing too much for the function
to track it properly
–200
Execution Error;
Store ref trace
before turning on
Normalize
The Reference trace data must be stored in the Ref trace before you turn on the
Normalization function
–200
Execution error; Sync
word was not found.
NADC & PDC: In an EVM measurement, the sync word is not found and the
synchronization cannot be established when Sync Word is selected in the Burst
Sync menu.
Flexible Digital Demodulation: The sync word cannot be detected because of
inappropriate parameter settings or incorrect signal
–200
Execution error;
Trace file contains
no compatible traces.
The trace file may have been created by another version of the Phase Noise
personality, which uses a different trace format that is incompatible with the
version you are running. Please check that you are running the most
up-to-date version of the personality
–200
Execution error;
Trace file created by
incompatible version
of Phase Noise App
The trace file may have been created by another version of the Phase Noise
personality, which uses a different trace format that is incompatible with the
version you are running. Please check that you are running the most
up-to-date version of the personality
–200
Function not
available before
Marker Zoom is
performed
Function not available before perform Marker Zoom
–200
Must perform Scan
before do Search
Cannot perform Search as the trace data is empty
–200
No marked signal
Cannot perform the selected function because no signal was marked. You must
mark the peak of interest before selecting the function
–200
No Measure At Marker
Added to Signal List
No signal peak was added into Signal List as there is no valid measure to
Marker result. You must perform Measure at Marker before selecting Measure
At Marker --> List
–200
No Peak Added to
Signal List
No signal peak was added from the Search to Signal List as no signal peaks
were found within the defined parameters of the search criteria
–200
Signal List is Empty
Cannot perform the selected function because the signal list contains no data
Err#
Message
Verbose/Correction Information
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...