Keysight N9038A MXE Service Guide 75
Instrument Messages
Advisory Messages
LO Fixed freq should be
greater than RF Stop freq
The setup frequencies break the rules for a downconverter measurement. The
measurement will still run, but check setup frequencies are correct before
continuing. The LO fixed freq should be greater than the RF Stop freq’s for an LSB
or DSB (for DSB measurements the setup uses LSB values) downconverter fixed LO
setup. Use the graph icon on the DUT setup form to clarify the setup required
LO Start freq should be
greater than IF Fixed
freq
The setup frequencies break the rules for an upconverter measurement. The
measurement will still run, but check setup frequencies are correct before
continuing. The LO Start freq should be greater than the IF fixed freq for an LSB
upconverter swept LO setup. Use the graph icon on the DUT setup form to clarify
the setup required
LO Start freq should be
greater than RF Start
freq
The setup frequencies break the rules for a downconverter measurement. The
measurement will still run, but check setup frequencies are correct before
continuing. The LO start freq should be greater than the RF Start freq’s for an LSB
downconverter swept LO setup. Use the graph icon on the DUT setup form to
clarify the setup required
LO Stop freq should be
greater than RF Stop freq
The setup frequencies break the rules for a downconverter measurement. The
measurement will still run, but check setup frequencies are correct before
continuing. The LO Stop freq should be greater than the RF Stop freq’s for a DSB
(for DSB measurements the setup uses LSB values) downconverter swept LO
setup. Use the graph icon on the DUT setup form to clarify the setup required
Low-pass filter set to
OFF
Turning on any band pass filter will turn off low-pass filters
No spurs have been found
You started a measurement in examine meas type in single or continual sweep
mode, or full meas type in single sweep mode, but no spurs were found
Preparing Calculation…
Preselector is centered
The preselector has been successfully centered
Preselector not used in
this frequency range.
You cannot center or adjust the preselector because it is not used at all at the
current marker frequency or between the current start and stop frequencies
Probe connected, cal data
is being reapplied;
<port>; <probe>
A probe has been connected, calibration data is being reapplied
Probe connected, no probe
cal; using cable cal
data; <port>; <probe>
A probe has been connected and no probe calibration data is available. The latest
cable calibration data will be used
Probe disconnected,
reverting to cable
calibration data; <port>
A probe has been disconnected, calibration data reverting to the last cable
calibration data
Reading SNS data…
The Keysight Smart Noise Source has been connected and the application is
reading the device EEPROM data. Please wait until complete before continuing
Recalled File <filename>
A file recall (open/load) was successfully completed
Message
Description/Correction Information
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...