Boot Up and Initialization
Potential Problems During Boot Process
Signal Level Verification
Frequencies Less Than 3.6 GHz
The basic functionality of the instrument below 3.6 GHz can be quickly verified
with the use of the internal 50 MHz calibrator signal. However, if there are
suspected problems at other frequencies an external signal generator will also
be needed:
1.
Switch the instrument into the Spectrum Analyzer mode by pressing
Mode
,
Spectrum Analyzer
.
2.
Turn on the internal 50 MHz Calibrator signal by pressing
Input / Output
,
RF Calibrator
,
50 MHz
.
3.
Tune the instrument to 50 MHz with a span of 10 MHz by pressing
Freq
,
50 MHz
,
Span
,
10 MHz
.
4.
Place the marker on the peak of the calibrator signal by pressing
Peak
Search
.
5.
Verify that the on screen of the 50 MHz Calibrator signal is
−
25 dBm
(+/-2 dB), as shown in
Figure 2-10
50 MHz Calibrator Signal
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...