170 Keysight N9038A MXE Service Guide
RF Preselector Section
RF Preselector Section Purpose
A22 Radiated Filter Assembly
The A22 Radiated Filter assembly provides the preselector filters that are
required in the frequency range of 30 MHz to 3.6 GHz.
There are a total of eight different signal paths through the A22 Radiated Filter
assembly which are broken up into two main sections. These two sections are
the Tunable Filters and the Fixed Filters.
Tunable Filters
The 30 MHz to 2 GHz frequency range of the tunable filters is broken up
into 6 different bands. They are:
— Band 0 - 30 MHz to 70 MHz
— Band 1 - 70 MHz to 150 MHz
— Band 2 - 150 MHz to 300 MHz
— Band 3 - 300 MHz to 600 MHz
— Band 4 - 600 MHz to 1000 MHz
— Band 5 - 1 GHz to 2 GHz
Bands 0 through 3 each have a lowpass prefilter to reject image signals that
are above their frequency range.
All of the tunable filters are adjusted at the factory and can be readjusted in
the field by use of the Service Menu. For detailed information on the Service
Menu see
Chapter 16, “Service and Diagnostics Menus”
The outputs of the six different tunable filter paths are separated into 2
different signal paths. Each of these two signal paths has its own limiter and
amplifier.
Fixed Filters
The fixed filter section covers a frequency range of 2 GHz to 3.6 GHz and
consists of two different signal paths. They are:
— Band 6 - 2 GHz to 3.6 GHz
— Through Path - 30 MHz to 3.6 GHz
The Band 6 filter path consists of a fixed highpass filter.
The through path has no filtering and is only used during the instrument
alignment and calibration.
There is a limiter and amplifier at the output of the fixed filter path section.
Refer to
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...