Keysight N9038A MXE Service Guide 85
Instrument Messages
Event Messages
–221
Settings conflict;
Feature not supported
for selected source
You have selected a feature that the selected source does not support
–221
Settings conflict;
Feature not supported
for this Input.
Some functionality is not available when certain Inputs are selected. For
example, Trigger Holdoff is not available for the BBIQ input
–221
Settings conflict;
Feature not supported
for this measurement.
Some functionality is available in one measurement, but not in another. (See
the measurements under the
Meas
key.) This error occurs if you send a
SCPI command or push a grayed-out key that is not available in the current
selected measurement
–221
Settings conflict;
Feature not supported
for this model number
This functionality is not a part of the instrument you are using, but may be
found in other models in the X-Series
–221
Settings conflict;
FFT IF Gain High not
available when Swept
IF Gain = Manual Low
When Swept IF Gain is manually set to Low, you cannot set the FFT IF Gain to
High because that would make the Reference Level couplings wrong in FFT
mode
–221
Settings conflict;
FFT method is
unavailable for level
gating
If you are using level gating, you cannot select the FFT Gate Method
–221
Settings conflict;
FFT sweep type is not
available while in
Gated LO
The gated LO function turns the LO on and off as it sweeps, so the FFT sweep
type is not available if you have selected gated LO
–221
Settings conflict;
FFT sweep type is not
available while in
Gated Video
The FFT sweep type is not available if you have selected the gated video
function
–221
Settings conflict;
FFT Sweeps
unavailable in
Tracking Source mode
Since FFTs do not sweep, you cannot use a Tracking Source while doing FFTs
–221
Settings conflict;
Fixed marker adjust
not available while
Marker Function is on
If a Marker Function is on for a Fixed marker, the marker's reported value is
derived from the function. Therefore, you cannot directly set the X or Y value
of a Fixed marker that has a marker function turned on
–221
Settings conflict;
Fixed Marker Y value
is not adjustable
with Normalize On
If Normalize is on the Amplitude scale is in dB units, so adjusting the Y value
of a Fixed marker is not possible
Err#
Message
Verbose/Correction Information
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...