244 Keysight N9038A MXE Service Guide
RF Downconverter Section
RF Downconverter Section Troubleshooting
will blink whenever the heater is turned on, which will be more frequent for
lower frequencies - when the current through the YTF main coil is lower. It will
continue blinking even when the instrument is tuned below 3.6 GHz, where the
YTF is not being used, so that it always remains warm.
Both DS4 and DS6 will turn on during the time that the instrument application
initializes the hardware, but this will happen so fast that it may only be possible
to see them change by measuring the voltage across them with an
oscilloscope. They will also be turned on periodically as the instrument is
running, but again for only a very brief period of time - tens of milliseconds.
Supply Voltage Verification
The A12 YTF Preselector can also have a problem initializing if the power
supply voltage to it is not correct or if the voltage regulator on the assembly is
faulty. To verify these voltages:
1.
View the status of the LED DS1 on the A12 YTF Preselector assembly, as
If DS1 is on the +30 volt DC supply to the assembly as well as the
+3.3 volt regulator are both fine.
If DS1 is not on and the instrument has either Option 508 or 526
proceed to Step 2. Otherwise, the cause for the low voltage level
will need to be determined. While this could be caused by a faulty
voltage regulator on the A12 YTF Preselector itself or by a faulty W4
YTF Control cable, it could also be a problem with the A15 Front End
Control assembly or the A6 Power Supply assembly. See the
associated chapters in this manual to verify the functionality of
these two other assemblies.
2.
(Option 503, 508, or 526 Only) Using a volt meter, check the status of the
+30 volt supply voltage to the A12 YTF Preselector assembly by
measuring from J3 to chassis ground, as shown in
If the voltage is b28 and +32 volts yet DS1 is not on the
most probable cause for the failure is the A12 YTF Preselector.
If the voltage is less than +28 volts the A12 YTF Preselector is most
likely not the source of the failure, the cause for the low voltage
level will need to be determined, which will most likely be caused by
either the A6 Power Supply assembly or the A15 Front End Control
assembly.
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...