Keysight N9038A MXE Service Guide 83
Instrument Messages
Event Messages
–221
Settings conflict;
Cal only available
when Source Mode is
Tracking
You must be in Tracking Source mode to use the Cal functions under
Normalize. Press Source, Source Mode and set it to Tracking
–221
Settings conflict;
Calibration cannot be
performed without
valid ENR data
The cal ENR table has no values in it, and hence the cal cannot be performed.
Correct by either populating the cal ENR table, set ENR mode to Spot, or set
the ‘Use Meas Table Data for Cal’ to ‘On’
–221
Settings conflict;
Cancellation is not
available while
measuring DANL floor.
Phase Noise cancellation does not make sense when measuring the DANL
Floor, so for this reason it has been disabled
–221
Settings conflict;
Cancellation Ref
trace has no data.
When performing phase noise cancellation, you need to supply a reference
trace that will be used to cancel out the background noise of the analyzer.
The reference trace must be in Reference (View) mode, and selected by the
Ref Trace parameter under the Cancellation menu
–221
Settings Conflict;
Cancellation trace
has different X-Scale
Reference trace for the cancellation has a different range of X-axis against
the target trace
–221
Settings conflict;
Cannot optimize while
user cal in progress
Optimize Preselector cannot be performed while a user cal is in progress. The
user cal performs an optimize preselector prior to taking the noise source
on/off level results for the cal data
–221
Settings conflict;
Can't Auto-Couple Res
BW in Zero Span
The resolution bandwidth cannot be set to auto while you are in zero span
(time domain)
–221
Settings conflict;
Carrier freq not
allowed with BMT.
(Bottom/Middle/Top
only)
The transmit band spur measurement only allows bottom (B), middle (M), and
top (T) channel frequencies for each supported frequency band. The carrier
frequency must be set to the bottom, middle or top frequency of the current
frequency band
–221
Settings Conflict;
Code channel
duplication
This error is reported when the given code channel overlaps other code
channel
–221
Settings conflict;
Continuous Peak is
not available with
Fixed marker
The continuous peak feature cannot be used with a marker that is fixed. By
definition that marker value cannot change
–221
Settings conflict;
Continuous Peak is
not available with
Signal Track on
The continuous peak feature cannot be used while you are also using the
signal tracking function
Err#
Message
Verbose/Correction Information
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...