116 Keysight N9038A MXE Service Guide
Instrument Messages
Condition Messages
Condition Errors 201 to 299, Signal Integrity
This series of errors corresponds to the bits in the
STATus:QUEStionable:INTegrity:SIGNal
sub-register. The second
column in the table shows the corresponding bit in that register.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be viewed in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
159
13
Settings Alert; Analog Out
settings conflict
W
The user has manually set the Analog Output
under the Input/Output menu to a setting that
conflicts with the current measurement.
There
will be no output on the Analog Out port until
this conflict is resolved.
In most cases, simply
set Analog Out to Auto for the optimal setting.
159
13
Settings Alert;I/Q
mismatch:<Differential|Inpu
t Z|Attenuation>
W
The impedance, differential, or attenuation
settings for the I and Q channels do not match.
For valid I+jQ measurements the impedance
and differential settings should be the same on
both channels and the attenuation should
match within 1 dB
159
13
Settings Alert;Parm/data
mismatch
W
For
Bluetooth
, the detected parameters did
not match the data
159
13
Settings Alert; Src pwr
ramp>ALC range
W
You have selected a Power Sweep range that
exceeds the ability of the external Source to
follow without changing mechanical
attenuation. Lower the Power Sweep range.
159
13
Settings Alert; Sweep Rate
Unavailable
W
The auto coupled sweep time exceeds the
maximum allowed. Therefore, full amplitude
accuracy cannot be attained. Please increase
the RBW or reduce the span.
159
13
Settings Alert; Span:RBW
Ratio too big
W
The chosen large ratio of span to RBW is not
possible. Please reduce the span or increase
the RBW and/or FFT Width.
161
14
Setting Modified
E
161
14
Setting Modified; Filter
not applied
E
The filter you have selected is larger than the
sampling frequency. You should select a
different filter.
a. The Input Overload error is not reported to the SCPI queue unless the :SYSTem:ERRor:OVERload ON
command has been issued, however it always sets the status bit. For details of the
:SYS
-
Tem:ERRor:OVERload
command, see the X-Series User’s & Programmer’s Reference.
Err#
Bit in
status
register
Message
Error or
Warning
More Information
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...