Keysight N9038A MXE Service Guide 99
Instrument Messages
Event Messages
–224
Illegal parameter
value; Illegal
identifier
<identifier>. This
value may already be
in use.
The value for the LXI LAN identifier parameter must be unique (that is, LAN0
and LAN7 must have different identifier strings)
–224
Illegal parameter
value; Index out of
range
When querying the LXI Event Log or the Servo Log, an index may be used to
look at a specific entry. This error occurs if the index provided does not point to
a valid entry
–224
Illegal parameter
value; Invalid list
length
You are trying to set some list measurement settings, but the multiple lists that
you sent had differing lengths. The number of settings must be consistent from
list to list
–224
Illegal parameter
value; LXI Event
<event> already
exists.
This error occurs when you try to add an LXI Event that has already been added
–224
Illegal parameter
value; LXI Event
<event> contains
illegal characters.
When a new LXI Event is created, it may not use the comma, semicolon, or
newline characters. All other printable ASCII characters are valid
–224
Illegal parameter
value; LXI Event
<event> does not
exist.
The requested event has not been added yet
–224
Illegal parameter
value; Measurement
not available
You tried to turn on a measurement that is not available in the current mode
–224
Illegal parameter
value;This instrument
is always DC coupled
You cannot set AC coupling in this instrument
–224
Illegal parameter
value;This model is
always AC coupled
You cannot set DC coupling in this analyzer
–225
Out of memory
There is not enough memory to perform the requested operation
–225
Out of memory;
Insufficient
resources to load
Mode <mode name>
If you attempt to load a mode via SCPI that will exceed memory capacity, the
Mode does not load and this message is returned. “mode name” is the SCPI
parameter for the Mode in question (for example, SA for Spectrum Analyzer
Mode.) You can free up resources using the
System, Power On,
Configure Applications
menu
–225
Out of memory; Memory
limit caused Data
Acquisition to be
truncated
Err#
Message
Verbose/Correction Information
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...