Keysight N9038A MXE Service Guide 113
Instrument Messages
Condition Messages
Condition Errors 101 to 199, Measurement Integrity
This series of errors corresponds to the bits in the
STATus:QUEStionable:INTegrity
register. The second column in the table
shows the corresponding bit in that register.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be viewed in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 141 indicates an Input Overload condition has been
detected, error 1129 indicates that failure has been cleared.
Two bits in this register are “summary bits” for registers at a lower level. There
are no error messages associated with these bits; they exist only as status bits,
which can be read with a
STATus:QUEStionable:INTegrity?
event query
or a
STATus:QUEStionable:INTegrity:CONDition?
query.
Note that these summary bits summarize the state and history of the event
registers at the lower level. This is true even for bits in the
STATus:QUEStionable:INTegrity
condition register. This means that:
— The summary bits read by the
STATus:QUEStionable:INTegrity:CONDition?
query are true if any
event bits are set in any of the :INTegrity sub-registers :SIGNal or
:UNCalibrated.
— The summary bits read by the
STATus:QUEStionable:INTegrity?
event
query are true if any event bit has undergone a false-to-true transition with
the PTRansition filter set, or a true-to-false transition with the NTRansition
filter set, in any of the :INTegrity sub-registers :SIGNal or :UNCalibrated
Thus, the summary bits
cannot
be used to determine the current state of a
lower level condition bit; only the state and history of the lower level event bits.
This register is itself summarized as bit 9 of the STATus:QUEStionable register,
as described in the section
“Condition Errors 601 to 699, Error Summaries” on
.
Err#
Bit in
status
register
Message
Error or
Warning
More Information
133
0
Signal Summary
status bit
only
This bit is the summary bit for the
STATus:QUEStionable:INTegrity:SIGNal
sub-register.
135
1
No Result
E
135
1
No Result;Turn on MCE
E
To calculate Timing and Phase results in the
Code Domain Power view of Mod Accuracy, the
"Multi Channel Estimator" must be set to ON.
Otherwise these results are invalid.
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...