Keysight N9038A MXE Service Guide 97
Instrument Messages
Event Messages
–200
Signal List is Full
Cannot perform the selected function because the signal list is full. Please
clear the list
–200
Signal Selected is
not in the Signal
List
The signal selected is not the list. You only can perform the operation on signal
that is already available in the list
–201
Invalid while in
local
The command cannot be executed while the instrument in Local control
–202
Settings lost due to
rtl
A “return to local” control was forced and some settings were lost as a result of
this
–203
Command protected
The command could not be executed because it is disabled. It was disabled by
licensing or password protection
–203
Command protected;
feature not licensed
The specified feature, for example “N9073A-TR2” is not licensed. The license
may have expired. You cannot use it until you obtain a license
–210
Trigger error
A trigger error has occurred, but the exact problem cannot be specifically
identified
–211
Trigger ignored
A GET, *TRG or other triggering signal was received, but was ignored because
of timing considerations. For example, maybe the instrument was not ready to
respond when the command was received
–212
Arm ignored
An arming signal was received, but it was ignored
–213
Init ignored
An initiate trigger/sweep request was received and ignored, because another
measurement was already in progress
–214
Trigger deadlock
The trigger source for the initiation of a measurement is set to GET, and the
following measurement query was received. The measurement cannot be
started until a GET is received, but the GET would cause an INTERRUPTED
error
–215
Arm deadlock
The arm source for the initiation of a measurement is set to GET and the
following measurement query is received. The measurement cannot be started
until a GET is received and the GET would cause an INTERRUPTED error
–220
Parameter error
A problem was found with a program data element. The exact problem cannot
be specifically identified
–221
Settings conflict;
There are many types of settings conflict errors. For full details, see
Settings Conflict Errors” on page 82
–222
Data out of range;
A data element was found but the instrument could not be set to that value
because it was outside the range defined for the command. A descriptive
message may be appended, such as “clipped to upper limit”
–222
Data out of Range;
clipped to source
max/min
A source parameter has been entered that exceeds the range of the selected
source. The parameter has been clipped to match the range of the source
Err#
Message
Verbose/Correction Information
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...