262 Keysight N9038A MXE Service Guide
L.O. Synthesizer
A14 L.O. Synthesizer Assembly Description
A14 L.O. Synthesizer Assembly Description
A14 L.O. Synthesizer Purpose
The A14 LO Synthesizer assembly has only one purpose and one output signal,
the 1st Local Oscillator (LO) for the A13 RF Front End assembly.
A14 L.O. Synthesizer Description
The main oscillator on the LO Synthesizer assembly is a Voltage Controlled
Oscillator (VCO) that has a frequency range from 570 MHz to 1150 MHz. This
VCO is then multiplied to the desired 1st LO frequency range required by the
A13 Front End assembly.
This VCO/Multiplier technology uses band pass filters after each multiplier
stage to suppress the undesired harmonics.
The LO Synthesizer assembly phase locks the 1st LO to the 4800 MHz
reference signal that is generated and phase locked to the 10 MHz reference
on the A16 Reference Assembly.
The quality of the 4800 MHz reference signal phase lock loop, the 1st LO phase
lock loop, as well as the 10 MHz internal reference signal all have a direct
bearing on the overall frequency stability of the instrument. Users can also
provide their own external reference signal to be used instead of the internal 10
MHz reference, which will also have a direct bearing on the overall stability of
the instrument.
After the different multiplier sections and amplifiers, the resultant frequency
range of the 1st LO provided to the A13 Front End assembly is 3.8 to 8.7 GHz
with an output power from +13 to +18 dBm.
The A14 LO Synthesizer assembly also contains a Field Programmable Gate
Array (FPGA) chip that is programmed from an onboard PROM chip when the
instrument is powered on. This code is updated if necessary when the
instrument software is updated.
Refer to
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...