370 Keysight N9038A MXE Service Guide
Power Supply & Midplane
A6 Power Supply Description
Control/Status Outputs
There are a number of control and status outputs for the A6 Power Supply
assembly. The ones that you will want to be familiar with are:
POWER_OK
is used to verify that the +12VD, +5.1VD, and +3.35VD are all on and within
specification. A TTL high level on this output brings the CPU out of reset and
initiates the instrument boot up process.
LINE_TRIGGER
is a TTL level signal that is synchronous to the AC line input. This signal is
typically used in spectrum analyzers to trigger an instrument sweep
synchronous to the AC power line.
FAULT_L
will be a TTL low level signal to indicate that the supply is experiencing an over
voltage, over current, or over temperature condition.
Fuse
The A6 Power Supply has no user replaceable fuse. While there is a fuse
internal to the supply this is not meant for field replacement. If the internal fuse
is blown, the power supply has experienced a major failure and should be
replaced.
Standby Supplies
The A6 Power Supply assembly has two standby supplies that should always
be on if the AC input voltage requirements are met. These are the +15VSB and
the +5.1VSB supplies. These supplies are used by the instrument to keep
certain circuits alive even when the power is turned off.
Over Current Protection
The A6 Power Supply assembly has built in over current protection that will
shut down the supply if current draw from the instrument is too great. The
power supply will remain on in over current state for a minimum of 1 second.
The power supply shall turn off no later than 5 seconds after the beginning of
the over current state. The power supply shall remain off until the line voltage
is removed and then reconnected or the front panel power switch is cycled.
Over current shut down does not apply to the standby supplies, the fan
voltage, and the +32VA supply.
Thermal Protection
The A6 Power Supply assembly will protect itself by shutting down if it
overheats. It will also reset itself with no user interaction after the temperature
is reduced by approximately 10 degrees C.
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...