Keysight N9038A MXE Service Guide 95
Instrument Messages
Event Messages
–200 to –299, Execution Errors
This section lists all messages in the range
except
–221 error messages. For
details of –221 messages, see
“–221 Settings Conflict Errors” on page 82
.
Note that Execution Errors are divided into subclasses, as follows:
–21x
Trigger errors
–22x
Parameter error
–23x
Data corrupt or stale (invalid data)
–24x
Hardware error
–25x
Mass storage error
–26x
Expression data error
–27x
Macro error
–28x
Program error (a downloaded program-related execution error)
–29x
Memory use error
Err#
Message
Verbose/Correction Information
–200
All ranges are off.
Turn on at least a
range
There are no range turn on in scan table. You need to turn on at least a range to
initiate a scan
–200
At Full Zoom
Marker Zoom is not available as it has reached full zoom
–200
At Full Zoom
Marker Zoom is not available as it has reached full zoom
–200
Execution Error
A program execution error has occurred. The exact problem cannot be
specifically identified
–200
Execution error;
Carrier frequency
outside device’s
transmit band
The entered channel/carrier frequency is not within the range of your current
mode setup selection of standard and device
–200
Execution error;
Invalid GSM burst
timing
A GSM-like burst was acquired, but its timing is not valid. Ensure the correct
Burst Type has been selected
–200
Execution error;
Invalid IP address
The IP address supplied is either not valid or does not belong to a compatible
Signal Generator. Please check the IP address and instrument connection and
try again
–200
Execution error;
Invalid Marker Trace.
Cannot place markers on the reference trace, because the reference trace is
currently turned off or has no data
–200
Execution Error; No
peak found.
No signal peak was found within the defined parameters of the search. (Note:
for ESA/PSA compatibility, the Swept SA measurement uses 780 for this error
number; all other measurements use –220)
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...