Keysight N9038A MXE Service Guide 713
Tests
Quasi-Peak Detector Accuracy
24.
Take a single sweep on the EMI receiver by pressing
Restart
.
25.
Once the sweep on the EMI receiver has completed and the auto peak
value is found record the marker delta amplitude value for the pulse
frequency being measured in
.
26.
Without changing any other setting on the EMI receiver or the IGUU 2916
EMI pulse generator change the Pulse Frequency on the pulse generator
and repeat
and
each.
Band B Testing
1.
Set the IGUU 2916 EMI pulse generator to output a 10 MHz Sine Wave
signal.
2.
Connect the IGUU 2916 EMI pulse generator Output Sine Wave Generator
to the EMI receiver RF Input.
3.
Preset the EMI receiver by pressing
Mode
Preset
.
4.
Tune the receiver to 10 MHz with a span of 0 Hz by pressing
FREQ
Channel
,
Center Freq
,
10 MHz
then
SPAN X Scale
,
Zero Span
.
5.
Set the Sweeptime to 2 seconds by pressing
Sweep / Control
,
2 s
.
6.
Set the RF Coupling to DC by pressing
Input / Output
,
RF Input
,
RF
Coupling DC
(instruments with Option 544 are always DC coupled so this
step is skipped for them.)
Table 22-29
Quasi-Peak Relative Response Band A - CISPR 16-1-1: 2010 Section 4.4.2
Pulse Frequency
(Hz)
Band A
Lower Limit
(dB)
Measured
(dB)
Upper Limit
(dB)
100
2.5
5.5
60
1.5
4.5
25
0
Reference
0
10
−
5.5
−
2.5
5
−
9.0
−
6.0
2
−
15.5
−
10.5
1
−
19.5
−
14.5
Isolated (0.1 Hz)
a
−
21.5
−
16.5
a. Due to the slow pulse rate you may need to take multiple sweeps to capture a complete
pulse.
Содержание N9038A
Страница 4: ...4 ...
Страница 16: ...16 Contents ...
Страница 66: ...66 Keysight N9038A MXE Service Guide Boot Up and Initialization Potential Problems During Boot Process ...
Страница 130: ...130 Keysight N9038A MXE Service Guide Instrument Messages Condition Messages ...
Страница 216: ...216 Keysight N9038A MXE Service Guide RF Preselector Section RF Preselector Section Troubleshooting ...
Страница 268: ...268 Keysight N9038A MXE Service Guide L O Synthesizer A14 L O Synthesizer Assembly Troubleshooting ...
Страница 354: ...354 Keysight N9038A MXE Service Guide Analog Digital IF A3 Digital I F Troubleshooting ...
Страница 406: ...406 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 407: ...Keysight N9038A MXE Service Guide 407 Block Diagrams Block Diagrams N9038A RF Input Option 503 508 526 Block Diagram ...
Страница 408: ...408 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams N9038A RF Input Option 544 Block Diagram ...
Страница 428: ...Keysight N9038A MXE Service Guide 447 Block Diagrams Block Diagrams LO Synthesizer Block Diagram ...
Страница 429: ...448 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams Reference Block Diagram ...
Страница 430: ...Keysight N9038A MXE Service Guide 449 Block Diagrams Block Diagrams MXE Computer Block Diagram ...
Страница 431: ...450 Keysight N9038A MXE Service Guide Block Diagrams Block Diagrams ...
Страница 436: ...Keysight N9038A MXE Service Guide 455 Service and Diagnostics Menus Service Menus Figure 16 5 Alignments Menu ...
Страница 500: ...Keysight N9038A MXE Service Guide 519 Replaceable Parts Hardware Figure 17 23 Front Frame Exploded View LED Backlight ...
Страница 504: ...Keysight N9038A MXE Service Guide 523 Replaceable Parts Hardware Figure 17 26 Front Frame Exploded View CCFL Backlight ...
Страница 589: ...608 Keysight N9038A MXE Service Guide Assembly Replacement Procedures Motherboard Assembly Figure 18 57 Bottom Screws ...
Страница 627: ...646 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 639: ...658 Keysight N9038A MXE Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 643: ...662 Keysight N9038A MXE Service Guide Instrument Software Software Updates ...
Страница 741: ...760 Keysight N9038A MXE Service Guide Tests Intermittent Unsteady Drifting Disturbances ...