Front Panel Menu Reference
SCPI Command
Display shows: “Enter measured data”
Measure the current with the Keysight 3458A and enter the data. Press Select
when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Steps 4 and 5 apply only to models N6761A and N6762A
Step 4. Disconnect the Keysight 3458A from the output terminals. Then select the second current
calibration point.
Front Panel Menu Reference
SCPI Command
The second calibration point is automatically selected. “Disconnect Meter”
should be displayed. Press Enter when done.
Step 5. Wait a minimum of 5 minutes for the internal temperature to stabilize. Then select the third
current calibration point.
Front Panel Menu Reference
SCPI Command
The third calibration point is automatically selected. “Wait 5 minutes, press
Next” should be displayed. Wait 5 minutes and press Next.
100 uA/200 uA Range Measurement - Models N6761A/N6762A with Option 1UA or 2UA
Step 1. Remove all loads from the output. Leave the sense terminals connected to the output
terminals.
Step 2. Select either the 100 μA measurement range or the 200 μA measurement range. Then wait 10
seconds.
Front Panel Menu Reference
SCPI Command
From the Calibration menu, select Current, 1UA/2UA Option Range Meas. Then
select Start.
or
CAL:CURR:MEAS 0.0002, (@1)
*OPC?
Step 3. Select the first current calibration point.
Front Panel Menu Reference
SCPI Command
The first calibration point is automatically selected. “Waiting 10 seconds”
should be displayed.
Step 4.Connect the Keysight 3458A and a 20 k
Ω
resistor to the output (see
).
Front Panel Menu Reference
SCPI Command
Measure the output current with the Keysight 3458A and enter the value. Press
Select when done. Press Next to finish calibration.
8 Verification and Calibration
396
Keysight N6705C Operating and Service Guide
Содержание N6705C
Страница 76: ......
Страница 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Страница 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Страница 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Страница 434: ......
Страница 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Страница 459: ...Overall Troubleshooting Keysight N6705C Operating and Service Guide 459 9 Service and Maintenance ...
Страница 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Страница 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Страница 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Страница 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Страница 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Страница 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Страница 476: ......